The Amplivox 270 Full Function Diagnostic Audiometer is a high-performance audiometer designed for comprehensive hearing tests. It meets international standards (IEC60601-1, IEC60601-1-2, IEC60645-1) and provides a range of output options, including air conduction, bone conduction, plug-in headphones, and sound field.
Key Features:
- Test Signals: Supports pure tone, impulse tone, warble tone, continuous sound, narrowband noise, and broadband noise.
- Frequency Range: Air conduction (125Hz–8kHz); Bone conduction (250Hz–8kHz).
-
Sound Intensity Range:
- Air conduction: -10 to 120 dBHL
- Bone conduction: -10 to 70 dBHL
- Sound field: up to 90 dBHL
- Accuracy: Sound output intensity is accurate within 3dB, with adjustable steps of 2.5dB or 5dB.
- Threshold Tests: Includes advanced functions such as Stenger, ABLB, and SISI tests.
- Control Features: A frequency adjustment button located near the intensity knob allows for one-handed operation.
- Display & Storage: LCD screen displays test settings and results, with the ability to store and print multiple audiograms.
- NOAH Software Compatibility: Easily integrates with NOAH software via a USB interface.
Included:
- Air conduction headset
- Bone conduction headset
- Patient transponder
- Audiogram
- Power cord
- Operation manual
- Carrying case
- Built-in microphone for operator-patient communication
This unit also features indicator lights for error-free operation and a built-in microphone for easy communication during tests. Additionally, the audiometer comes in a compact and lightweight design, making it portable and user-friendly for clinical or on-the-go diagnostics.
The Amplivox 270 is a two-channel diagnostic audiometer designed for comprehensive diagnostic audiology and hearing aid fitting. It includes capabilities for:
- Air conduction (AC)
- Bone conduction (BC)
- Speech testing
This device is suitable for use in clinics, workplaces, or even in mobile setups by audiology specialists and hearing health experts.
Test Options:
The audiometer provides several key testing features for both screening and diagnostic purposes:
- Manual air and bone conduction testing
- Speech testing (live or recorded): Provides SRT (Speech Recognition Threshold) and SDT (Speech Detection Threshold) scores.
- Clinical tests: Includes Stenger, ABLB (Alternate Binaural Loudness Balance), and SISI (Short Increment Sensitivity Index) for in-depth patient evaluation.
Design and Usability:
- Display: Clear, precise, and positioned at an optimal viewing angle.
- Key layout: Logical and user-friendly, with quick access to functions for reducing test times.
- Controls: Silent rotary controls to minimize noise and improve test accuracy.
Data Management:
- Internal memory: Allows test results to be saved or printed immediately.
- Amplisuite software: Used for easy downloading, managing, and processing of results, with additional counseling tools like speech bananas.
- Third-party connectivity: Integration with EMR systems such as NOAH and OtoAccess® for seamless workflow.
Key Features:
- Channels: Two-channel audiometry
- Frequency range (AC): 125Hz - 8000Hz
- Attenuation range (AC): -10dBHL to 120dBHL
- Special tests: ABLB, Stenger, and SISI
- Connectivity: EMR integration with NOAH and OtoAccess®
Technical Specifications:
- Air Conduction Frequency Range: 0.125kHz, 0.25kHz, 0.5kHz, 0.75kHz, 1kHz, 1.5kHz, 2kHz, 3kHz, 4kHz, 6kHz, 8kHz
- Bone Conduction Frequency Range: 0.25kHz, 0.5kHz, 0.75kHz, 1kHz, 1.5kHz, 2kHz, 3kHz, 4kHz, 6kHz, 8kHz
- Frequency Accuracy: <1%
- Distortion: <2%
- Tone Presentation: Continuous, interrupted, pulsed, warble
- Power Supply: 240Vac 50/60Hz
- Physical Dimensions: 355mm (L) x 230mm (W) x 50mm (H)
- Weight: 1540g / 3.39lbs
Standard and Optional Equipment:
-
Standard equipment:
- Audiometric headset
- Bone conduction headset
- Patient response switch
- Power supply with adaptors
- Carry case
- Audiogram cards (50)
- Manual and software (available via download)
-
Optional equipment:
- Amplivox Audiocups
- Masking earpiece
- Insert earphones
- Thermal printer
- OtoAccess® database
- Combined microphone and monitor headset
- Talk-back microphone
- Loudspeaker kit
- USB cable (PC connection)
- Booth leads
- Power bank USB cable
Data Management and Masking:
- Internal database: Stores up to 12 audiograms
- Data transfer: Via USB to Amplisuite, NOAH, or OtoAccess®
- Masking: Narrowband and speech-weighted masking
- Masking output: 90dBHL max for frequencies 250Hz - 4kHz
Speech and Free Field Capabilities:
- Speech input: Live speech, MP3, or CD input
- Free field output level range: Up to 90dB
Compliance and Standards:
- Safety standards: Complies with IEC 60601-1 (medical electrical equipment safety), including UL, CSA, and EN deviations.
- Performance standards: IEC 60645-1/ANSI S3.6
- EMC: Compliant with IEC 60601-1-2