Description
Report Item-
Intelligent Device System:
The intuitive design and user-friendly setup of the MSL Electrosurgical Generator (diathermy) for use in the modern operating room. It features both monopolar and bipolar functions to satisfy all surgical demands with safety, flexibility, reliability, and convenience. -
REM (Return Electrode Monitoring):
The REM system monitors return electrode contact quality. It deactivates the generator if a fault in the patient/return electrode contact is detected, minimizing the danger of burning incidents. -
Compatible with Other Equipment:
- Argon Plasma Coagulation System
- Optimum Smoke Evacuation System
-
Multilanguage Available:
Language options include English, Spanish, Portuguese, Turkish, etc. -
Low Frequency Leakage:
- Normal Polarity, Intact Chassis Ground: <10μA
- Normal Polarity, Ground Open: <50μA
- Reverse Polarity, Ground Open: <50μA
- Sink Current, 140V Applied, All inputs <50μA
-
Application of Use:
- General Surgery
- Gastroenterology
- Dermatology
- Vascular Surgery
- Obstetrics and Gynecology
- Heart/Thoracic Surgery
- ORL/ENT
- Minimally Invasive Surgery (MSI)
- Cerebral Surgery
- Neurosurgery
- Orthopedics & Plastic Surgery
- Trans Urethral Resection (TUR)
-
Technical Specifications:
- Main Frequency: 512 KHz
- Power Supply: 100-240V / 50-60Hz
-
Weight and Dimension:
- Weight: 8.00KG
- Dimension: 341647
-
High Frequency Leakage:
- Monopolar: <150mA
- Bipolar: <60mA
-
Working Mode Maximum P-P Voltage, Rated Load, Max Output Power, Crest Factor:
- Monopolar Cut:
- Pure: 1900V, 500Ω, 300W, Crest Factor 1.6
- Blend: 2300V, 500Ω, 250W, Crest Factor 2.5
- Monopolar Coag:
- Fulgurate: 2800V, 500Ω, 120W, Crest Factor 2.9
- Spray: 7000V, 500Ω, 100W, Crest Factor 6.2
- Bipolar Coag Standard: 550V, 100Ω, 70W, Crest Factor 1.6
- Monopolar Cut: